Page 12 - Catalog-21
P. 12

Digital Force & Torque Gauges

    Specifications - Force Gauges

                                                                    Series 7                                             Series 5                                            Series 4
                                                               Professional                                            Advanced                                         Performance

    Accuracy (% of full scale)                                                       ±0.1%                                                                     ±0.2%
    Sampling rate
    Resolution                                          14,000 Hz                                            7,000 Hz                                          3,000 Hz
    Outputs
                                                                                     1/5,000                                                                   1/2,000
        RS-232
         Mitutoyo (Digimatic)                                                                 USB, RS-232, Mitutoyo, Analog
         Analog
         General purpose I/O                                                                  Configurable up to 115,200 baud
         Set point
    MESURTM Lite software included                                                   Serial BCD suitable for all Mitutoyo SPC-compatible devices

    Data output functionality                                      ±1 VDC, ±0.25% FS at capacity. Positive for compression, negative for tension

    ACSII command set for external control                                                    Three open drain outputs, one input
    Communication w/MESURTMgauge and MESURTM Lite
                                                                                                           Three open drain lines
    Communication with ESM301 test stand and Series DC
    controllers                                                                                              Yes

    Data memory                                         Automatic output or by pressing DATA button, or requesting from a PC or other device.  By pressing DATA button, or requesting from
    Data memory deletion                                                                                                                         a PC or other device (currently displayed
    Statistic calculations and output to PC                                                                                                                    reading only).
    Break Detection
    1st / 2nd Peak Detection                                                         Full command set                                                 “?” only (request current reading)
    Programmable footswitch command string
    Averaging mode                                                                   Full communication                                        Single point or continuous data output only
    External trigger mode
    Analog load bar graph shown on display                                           Full communication,                                       Stop at force set points, overload protection,
    Set points (for pass / fail indication)                                          including PC control                                       and communication through “?” character
    Password Protection
    Peak readings always displayed                                                                                                                                  only
    Programmable audible tones
    Calibration procedure                               5,000 readings, with output                        1,000 readings, with output                         50 readings, with output
    Overload outputs
                                                                                              Any individual data point, or clear all

                                                                                     Max, min, mean, std. dev., number of readings, with output

                                                        Yes* No

                                                        Yes* No

                                                        Yes No

                                                        Yes* Yes                                                                                               No

                                                        Yes**                                                Yes                                               No

                                                                                                             Yes

                                                                                     Yes with high / low indicators, markers on analog bar graph, and outputs

                                                                   Yes, for calibration and individual keys                                                    No

                                                                                                             Yes

                                                                                                             Yes

                                                                                     Configurable number of data points (up to 10 in each direction)

                                                                                     Dedicated pins for tension and compression

    Battery life                                        Rechargeable. Backlight on: up to 7 hours of continuous use / Backlight off: up to 24 hours of continuous use

    Units of measurement                                                             lbF, ozF, gF, kgF, N, kN, mN (depending on model)
    Safe overload
    Weight (lb [kg])                                                                 150% of full scale (display shows warning at 110% and above)

    Included items                                      M7-012 - M7-100: 1.0 [0.45]            M5-012 - M5-100: 1.0 [0.45]                            M4-012 - M4-100: 1.0 [0.45]
                                                        M7-200 - M7-500: 1.2 [0.54]            M5-200 - M5-500: 1.2 [0.54]                            M4-200 - M4-500: 1.2 [0.54]
                                                                                              M5-1000 - M5-2000: 1.8 [0.82]

                                                        Carrying case, chisel, cone, V-groove, hook, flat, extension rod, universal voltage AC adapter/charger, rechargeable battery, quick-start guide, USB
                                                         cable, resource CD (USB driver, MESUR Lite software, MESURgauge DEMO software, and user’s guide), NIST-traceable certificate of calibration

                                                                                                                                 with data

    * With configurable automatic data output, data storage, and zero function upon event completion. Can also stop travel of certain motorized test stands.
    ** Can stop certain motorized test stands upon event completion.

12